Biochrom’s Libra S4+ visible spectrophotometer is ideal instrument for education and QC laboratories. It is compact, lightweight and ergonomically designed for simpler applications. There is a large display for ease of reading, simple user interface for rapid set up and analysis.
It measures absorbance, % transmission and concentration as well as simple kinetics. The large display allows the user to view results in numerical value as well as a graphical view. The user can view the peak, zoom in and out all on the unit.
Up to 72 methods can be stored providing the user a quick and easy way to recall frequently used methods. These methods can also be locked using a pin protected password, as well as unlocked and deleted.
The S4+ provides the ability to label your sample for easy identification. Your sample can be measured in a test tube using the optional test tube cell holder accommodating test tubes that are 10 to 18 mm in diameter. A heated cell holder is also available for thermostatted measurements at 37°C.
Data Output
PVC (Print Via Computer) software is included with the instrument providing the user with the option to transfer data directly to a PC. Data can be easily stored on a USB memory stick and analyzed at a later date
Features
- Single beam visible spectrophotometer covering 325 – 1100 nm
- Save data on a USB memory stick and analyze results later
- Test tube and heated cell holder accessories are available
- PVC (Print Via Computer) software included to allow data transfer to PC from the instrument
Technical Specifications
| Specification |
Data |
| Type |
Visible |
| Wavelength Range |
325 – 1100 nm |
| Wavelength Accuracy (+/-) |
2 nm |
| Wavelength Reproducibility (+/-) |
1 nm |
| Lamp Source |
Pulsed Tungsten Halogen |
| Beam |
Single Beam |
| Optical System |
Monochromator |
| Bandwidth |
<7 nm |
| Detector |
CMOS Array |
| Photometric Range |
-0.300 to 2.500 A |
| Stray Light |
<1% T at 340 nm |
| Zero Stability |
±005A/hour near 0A |
| Baseline Flatness |
n/a |
| Measurement Type |
Fixed Wavelength, Abs/% T, concentration (factor), kinetics* |
| Cell Compatibility |
Cuvette 10 mm |
| Other Sample Containers |
Test Tube cell holder (10-18 mm)* |
| Cell Holder |
Single Cell |
| Accessory Options |
Heated cell holder* |
| Method Storage |
72 with PIN number protection |
| Compliance |
CE |
| Data Output |
Export to PC with USB cable, USB memory stick |
| Display |
LCD |
| Data Interface |
USB cable |
| Languages |
English, French, German, Italian, Spanish, Chinese and Japanese |
| Power Requirement |
90-265 V AC, 50/60 Hz, 15 VA |
| Weight |
1.6 Kg (3.5 lb) |
| Dimensions (WxDxH) |
32.5 x 22.5 x 13.3 cm (13 x 9 x 5 inches) |
| Noise |
<0.002 A RMS at 0 A and <0.003 RMS at 1 A at 600 nm |
| Photometric Accuracy |
<0.002 A at 0 A at 546 nm |
| Photometric Reproducibility |
<0.002 A at 0 A and 500 nm |
| Spectral Scanning |
N/A |